The 897A Series GM Detectors are designed to detect ionizing radiation utilizing the pulse conversion technique.
Key Features
Features 3 ranges available. Each covers 5 decades from 10-2 to 105 mR/h
Energy response: + 15% for 80 keV to 1.5 MeV
Preamp integral with detector
Seismically qualified
Life expectancy: approximately 105 rads
All power supplied from ratemeter
Ratemeter may be remotely located up to 2000 feet
Single cable between ratemeter/detector
8 µCi 36Cl check source
Although all the Model 897A Series GM Detectors are functionally identical, the detectors differ in the tube type employed, sensitivity, and housing. The GM detector has preamplifier circuitry located in the interior of the overall protective housing. The preamplifier provides pulse conditioning and cable driving capabilities to match the input characteristics of the ratemeter.
Rugged, Reliable Radiation Detector
Personnel Protection
Factory NIST Traceable Calibration
Field Proven Design
For Use with Model 956A Digital Readout or Model 960 Series Digital Radiation Monitoring System (DRMS)
Area monitoring is used for the detection of x-ray or gamma radiation in a selected area. The monitor should be used in any location where personnel may be exposed to an adverse amount of radiation. Applications include nuclear reactors, accelerators, hot cells, irradiators, and any area where radiation sources are handled. These monitors can be used as single channel monitors or grouped together as a multichannel area monitoring system.