with mechanical blocker
with electromagnetic blocker
with electric lock
The pass-through cabinet, installed directly into a partition, is used to transfer small objects and materials between rooms with different classes of sterility. The walls, the top and the bottom of the pass-through cabinet are made of sandwich panels. If laminar flow is desirable inside the cabinet, a HEPA filter is fitted in the opening on the top of the cabinet. To exhaust air from the cabinet, the right door is fitted with a grille at the bottom.
For the case when it is possible to install a regulating ventilation grille in the panel, the bottom of the cabin consists of a perforated metal plate, below which there is a sump tray.
The product is certified in EU countries.
The standard cabin layout is designed for a maximum load of 10kg.
The requirement for greater carrying capacity of the pass-trough cabin must be specified in the note.
Airflow
active
passive
Active - has its own intake of filtered air through a HEPA filter; air is exhausted to an area with a lower class of cleanliness.Passive - no HVAC treatment.
Blocker
electronical solenoid lock ST-DB100 with indication
The cabin has a mechanical blocker.
Cabin is not blocked
Cabin with indication