Working environment
a) Suction pressure: 0.1MPa (Absolute pressure)
b) Suction temperature: 20℃
c) Exhaust pressure: Table 1
d) Motor speed: Table 1
Working conditions
a) Ambient temperature: 5℃~ 40℃
b) Relative humidity: ≤80%
c) Voltage: 198V-242V
d) Frequency: 49Hz-51Hz
e) The ground where the compressor is placed should be flat, with the angle of inclination of the body after installation being no more than 2°.
When the pressure reaches 0.8MPa ±0.2 MPa, the compressor will automatically halt operation.
When the pressure is 0.5 MPa ±0.05 MPa, the compressor will automatically start.
A complete set of accessories is provided with the compressor, including:
a) Drive motor and necessary electrical starting equipment
b) Gas tanks, safety valves, pressure gauges
c) Required wearing parts and other accessories that need to be replaced during the warranty period
Electrical safety performance
The electrical safety performance complies with GB 9706.1-2007 "General requirements for medical electrical equipment Part 1" standards.