Beyond Heat
The technology that overcomes the limitations of traditional tecar therapy
The Technology
The effect of biostimulation of tissues, fundamental for the therapeutic efficacy of diathermy, is not only linked to the development of heat, but also to the exposure of tissues to an electromagnetic field capable of interacting with cells and their metabolism.
For this to happen, a voltage higher than 100 V is required which, in order to cross the cell membrane, must be modulated at frequencies of the order of hundreds of KHz.
With traditional devices for tecartherapy (which regulate power through voltage regulation), when this value of about 100 V is reached, overheating of tissues develops simultaneously and is often not tolerated by patients.
eMOTION TECAR overcomes this limit thanks to a particular modulation of the electromagnetic field, succeeding in developing the maximum biostimulation of the tissues without generating an annoying heat for the patients.
PWM (Pulsed Width Modulation) Technology
The generated signal is a sine wave with frequency 450 KHz / 680KHz and always maximum voltage. This wave is modulated with a square wave with a constant repetition frequency and a variable duration.
The power delivered is precisely regulated by means of the Duty Cycle (TON / T). This technology drastically reduces the dissipated power, thus allowing the generation of heat to be maintained at a level that is tolerable for the patient but without limiting the power supplied.