Inputs: 5 or 6 analog inputs (transducers), 10/12 digital (switches)
Outputs:
10 or 12 outputs digital signals used for: a) repeating to second panel, b) central alarm system, c) BMS of installation.
built in 10 or 18 LED, 2 red LED for each gas, 1 for high and 1 for low pressure indication and 1 green for Normal
built in buzzer for alarm signaling
LCD display (2×16 characters) for real time pressure monitoring and text messages corresponding to alarm conditions
adjustable limits and alarm messages (by user programming through TCP/IP web page interface)
adjustable analog sensors range and type
alarm programming through Ethernet connection with an IP address
button test and silence
230V 50Hz-60Hz power supply (double insulated transformers)
alarm / events report and log file (max 200 records with relative time stamp)
editable gas names
remote control and monitor of the central AGSS central units through the L6
leakage test procedures according to EN ISO 7396-1