MC150 - STABILIZER is a second stage pressure reduction unit with the task to equalize the eventual pressure variation in the hospital pipeline system to ensure a correct pressure from the terminal units.
MC150 - STABILIZER is a second stage reduction unit where the primary gas supply is provided by high pressure gas manifolds (such as MC80, MM90 or liquid oxygen tank (LOX)). When there is a signal for pressure deviation in relation to the alarm settings, the alarm can easily be displayed on a Gas alarm unit. It is also possible to send information to the central operation control. The stabilizer can be delivered with either pressure transmitter 4-20 mA or with contact gauge.
The gas reduction unit MC150 - STABILIZER must always be installed in compliance with the standards EN ISO 7396-1 and the appropriate national standards.
FOR A COMPLETE MC150 STABILIZER MANIFOLD ADD:
Alarm unit (included if ordered together with HP unit)
TECHNICAL DATA
Gases:O2, Air, N2, N2O, CO2, (all medical gases)
Nominal flow:360 m3/h
Inlet maximal pressure:20 bar (2000 kPa)
Outlet nominal pressure:4,5 bar (setting range 0,5–6 bar)
Inlet connection:2× G1 1/2"F + soldering piece pipe ∅ 35 mm
Outlet connection:2× G1 1/2"F + soldering piece pipe ∅ 35 mm
Pressure relieve valve:6,8 bar
Pressure relieve valve
pipe dimension: ∅ 35 mm
Regulatory status:
Complies with Medical Devices Directive 93/42/EEC
Complies with EN ISO 7396-1
(Central Gas Supply Systems)
Complies with EN 60601-1-2
(Electromagnetic compatibility)
present SIS HB 370 and HTM 02-01