- Individual sweep speed setting for each parameter
- 128 full hours trend data stored
- SpO2 measurement with very low perfusion detection and pulse beep
- 2 traces ECG
- screen freeze on/off
- connection to Central monitoring system using LAN or W-LAN
- High-performance ECG detection and analysis
• input impedance for dry skin or electrode patch
• ST level, PVC count real-time display
• 3 kinds of Arrhythmia analysis
• pacemaker detection function
- Enhanced data management
• graphical and tabular trends
• 20 cases per 10 sec alarm waveform stored, reviewed and printed
- Internal software selectable in GB, DE, FR, IT, PT, PL, RO, HU, ES, CZ, BG, RU, TR, CN
- User manual: GB, IT (on request: ES, FR, DE)
Technical Specifications
ECG
• Lead type: 3-lead, 5-lead (option)
• Lead selection:
- 3-lead I, II, III
- 5-lead I, II, III, aVR, aVL, aVF, V (option)
• Heart rate range:
- Adult: 30 to 300 bpm,
- Neonate/Pediatric: 30 to 350 bpm
• Heart rate accuracy: ± 1bpm or ± 1%, whichever is greater
• Sweep speed: 6.25, 12.5, 25, 50 mm/sec
• Filter:
- Diagnosis: 0.05 to 150 Hz
- Monitoring: 0.5 to 40 Hz
- Moderate: 0.5 to 25 Hz
- Maximum: 5 to 25 Hz
• S-T segment detection range: -2.0 to 2.0 mV
• Arrhythmia analysis: ASYSTOLE, VTACH, VFIB
• Pacemaker detection mode: Indicator on waveform display (user selectable)
• Protection: Against electrosurgical interference and defibrillation
NIBP
• Technique: Oscillometric
• Measurement Modes: Manual/automatic/continuous
• Pressure range: 0 to 300 mmHg ±3 mmHg
• Measurement range:
- Adult pressure: 20 to 260 mmHg
- Pediatric pressure: 20 to 230 mmHg
- Neonatal pressure: 20 to 120 mmHg