RTQC is a software package dedicated to radiotherapy departments, enabling them to maintain high standards of quality and safety. This module facilitates the acquisition and analysis of images for internal quality control, guaranteeing regular and reliable assessment of radiotherapy equipment performance. The module ensures that treatment devices meet the latest safety standards and comply with legal requirements.
Specifications
- Calibration: Automatic detection and evaluation with defined tolerances for pixel deviation.
- Uniformity and Noise: Automatic analysis with 5% VOC tolerance. - Dose Stability: Comparison of measured and reference profiles, 2% tolerance.
- Spatial Distortion: Quadrant-by-quadrant measurement with 1 mm tolerance.
- Centering: Isocenter and collimator evaluation, 2 mm tolerance.
- Picket Fence: Analysis with tolerances of 0.6 mm (static) and 1.2 mm (dynamic).
- Contrast Resolution: Evaluation via test images.
- CBCT analysis: Image quality, resolution, contrast, slice thickness.
Added value :
- Easy to use: Intuitive interface and clear navigation.
- Automated reporting: Monthly and half-yearly PDFs.
- Extensive coverage: Compatible with various radiotherapy equipment.
- Compliance: Compliance with ASN and ANSM standards.
- Interoperability: RIS/PACS integration, DICOM support.
- Real-time monitoring: Control to ensure patient safety and avoid interruptions.