4-pole electrode in carbon plastic technology
Galvanic insulated RS485 Interface
Different modes for conductivity temperature compensation
Multifunctional
The electrode is multifunctional and may be used as a flow through electrode as well as a dip electrode.
Conductivity Measurement
We use modern state of the art technology 4-pole electrode, utilising carbon compound technology for conductivity measurement. This technology gives no polarization effects and results in an absolutely linear and stable function over years of use.
Temperature Compensation
Conductivity temperature coefficient can be adjusted, ensuring a high accuracy conductivity measurement. The sensor offers four modes for compensation. For the use in calibration of hemodialysis machines the Biotech-Lab software allows to select the machine temperature compensation value by brand name.
Temperature Compensation Modes
No compensation
Linear with one adjustable value
Dynamic with two adjustable values and temperatures
Non Linear Function for tap, purified and waste water in accordance with ISO 7888 – DIN EN 27888 Standards.
Calibration
PTB/NIST traceable
Communication rate
All values five times per second
Degree of protection
IP65
Power supply
3,3 to 5 V, max. 80 mA via RS485- or USB-Interface
Environmental conditions
Ambient Operating Temperature Range: 15 to 35°C
Storage Temperature Range: -10 to 60°C
Operating and Storage Humidity: 85% max
relative humidity, non-condensing
Dimensions -
124 x 34 mm