Integrated filter in the flow path in order to protect the sensor
The sensor comes equipped with a Quick-disconnect coupling
Other sensors can be adapted easily.
Measuring principal:Turbine
Range:100 to 2000 ml/min H2O at 22°C / 71.6°F
Resolution:1 ml/min
Accuracy:± 2.0 % of reading
Fluid temperature:-20°C to +80°C
Calibration:PTB/NIST traceable
Communication rate:Five times per second
Degree of protection:IP65
Power supply:3,3 to 5 V, 60 mA via RS485- or USB-Interface
Environmental conditions:Ambient Operating Temperature Range: 15 to 35°C
Storage Temperature Range: -10 to 60°C Operating and Storage Humidity: 85% max relative humidity, non-condensing
Dimensions:110 x 30 x 30 mm