Ensure uptime with accurate, low-level sodium measurements and predictive diagnostics.
Benefits
Optimise Operation and Response Time with Automatic Electrode Reactivation
To maintain optimum response time and accuracy, the NA5600sc analyser provides automatic electrode reactivation. Reactivation uses non-hazardous chemicals and eliminates the need for manual reactivation or electrode etching.
Space-Saving Design
Smaller instrument footprint with streamlined layout to allow for easy integration into existing or new sites.
Low Maintenance
Maintenance of the NA5600sc Sodium Analyser requires reagent replenishment only every 90 days and annual replacement of reagent tubing and the sodium electrode. Clear step-by-step instructions are provided to simplify maintenance operations.
Avoid Downtime
Predictive diagnostic tools, including Hach’s proprietary Prognosys technology, warning LEDs, and high visibility notification screens let you avoid unplanned downtime.
Specifications
Technical Attributes
Accuracy - 0.01 - 2 ppb: ± 0.1 ppb
2 - 10,000 ppb: ± 5%
Acidity - < 50 ppm
Ambient temperature - 5 - 50 °C
Analogue outputs - 6 isolated, 0 - 20 mA or 4 - 20 mA; load impedance: 600 Ohm maximum
Connection: 0.644 - 1.29 mm² (24 - 16 AWG) wire; 0.644 - 0.812 mm² (24 - 20 AWG) recommended, twisted pair shielded wire
Calibration method - Automatic with known addition
Manual: 1 or 2 points
Detection limit - 0.01 ppb
Dimensions - 681 mm x 452 mm x 335 mm (H x W x D)
Display - Coloured 5.7” LCD
Electrode type - Sodium ISE (ion specific electrode) electrode and reference electrode with KCl electrolyte
Flow rate - 100 - 150 mL/min (6 - 9 L/h)
Fuse - Input power: T 1.6 A, 250 VAC