Combines LipiGuard filter media technology with 40 μm lock woven microaggregate screen
Designed for use with either intra- or post-operatively salvaged blood that may be washed or unwashed, thereby reducing inventory requirements for a number of different filter products
Low volume (20 mL) housing allows excellent blood product recovery
High flow rate for rapid infusion
Easy spike insertion with a secure hold and rapid filter priming
Effectively replaces conventional microaggregate filters
Latex free
Performance Characteristics
Fat, leucocyte and C3a (and its immunosuppressive metabolite C3a des
Arg77) were studied for both washed and unwashed salvaged blood, obtained
intra-operatively and post-operatively, for the following procedures:
total hip replacement
coronary artery bypass graft
knee arthroplasty
prostatectomy
discectomy
ortho-laminectomy
femoral osteotomy
No statistically significant differences were found for the filter performance using either washed or unwashed, intra- or post-operatively collected blood.
Combined performance data are summarized in Figure 1.
Combined Performance Data
*Pre-filtration fat concentrations (globules/mL) averaged 21,452, 2,081 and 542 for ranges 40 μm, respectively. Fat challenge (volume) was calculated as the product of fat concentration (globules/mL) and volume of blood product (mL). The total volume of fat for each of the three ranges of fat globule sizes was calculated from estimates using the diameters 5, 25 and 60 μm for ranges of fat sizes (40 μm) measured, respectively. Data on file.