The Hamilton OneFerm pH sensor is a single use glass electrode for pH measurement. It retains the high accuracy performance of a glass electrode after gamma irradiation and dry storage. It is delivered as an integrated part of a single use container to provide ease of use. The OneFerm pH sensor is only available to integrators such as bag-, bioreactor- or downstream skid manufacturers.
Specifications
a-length - 425 mm
Electrical Connector - VP6
Sensor Output - Traditional: mV
Measurement Principle - Combination electrode; pH potential measured against reference
Measuring Range - pH 3 to 10Additional specifications
Sensitivity - Before gamma irradiation: 57 to 60 mV/pH at 25 °C (97 % -101 %); Shift after gamma irradiation: ± 1.5 mV/pH
Zero Point - Before gamma irradiation: ± 20 mV; Shift after gamma irradiation: ± 10 mV
Calibration Data - Yes, provided on sensor label
pH Accuracy - After gamma irradiation and dry storage: ± 0.15 pH with provided
calibration values. ± 0.1 pH after product calibration at the measurement temperature and within 2 pH units from the product calibration point
Drift at Room Temperature - < 0.1 pH per week after wet-in time, in buffer pH 7
Membrane Shape - Spherical
Diaphragm - HP-Coatramic
Temperature Sensor - Pt100
Diameter - 12 mm
Process Connection - PG13,5
Wetted Parts - Glass
VMQ (Silicone elastomer)
O-ring Material - FDA 21 CFR 177.2600; EG 1935/2004; USP ; USP Class VI (121 °C)
Biocompatibility - Yes, tested according to ISO 10993-5 and USP
Wet-in Time - 30 min after dry storage
Shelf Life - 12 months prior gamma irradiation. 18 months after gamma irradiation
Autoclavable - No
CIP - No
Steam Sterilizable - No