The VisiFerm DO SU ARC 225 C for single-use applications works with Hamilton single-use ODO Cap . The VisiFerm DO SU ARC 225 C is reusable and can be mounted directly into the ODO Cap within the bag or SU bioreactor.
Specifications
a-length - 215 mm
Electrical Connector - VP8
Sensor Output - Arc: Modbus, 4 to 20 mA
Measurement Principle - Oxygen dependent luminescence quenching, temperature compensation
Measuring Range - 4 ppb to 25 ppm (DO)
0 to 62.85 %-vol
0 to 300 %-sat
Accuracy at 25 °C - ± 1.0 %-sat (at 20 %-sat), only in combination with the gamma
irradiated Hamilton single-use ODO Cap and with product calibration
± 1.0 %-sat (at 100 %-sat), only in combination with the gamma irradiated Hamilton single-use ODO Cap and with product calibration
± 2.5 %-sat (at 200 %-sat), only in combination with the gamma irradiated Hamilton single-use ODO Cap and with product calibrationAdditional specifications
Drift at Room Temperature - < 1 % per week (of measurement value)
Electrolyte - None
Oxygen Consumption - None
Temperature Sensor - NTC 22 kOhm
Configurable Values - DO: mbar; %-sat; %-vol; µg/l; mg/l; ppb/ppm (gas); ppb/ppm (dissolved oxygen); Temperature: °C
Diameter - 12 mm
Analog Interface 1 - 4 to 20 mA for DO, programmable, galvanically not isolated; pulse width modulation 5 kHz
Digital Interface - RS485 Modbus (max. 31 addresses)
Baud Rate - 4800, 9600, 19200, 38400, 57600,115200 Bd
Operating Voltage - 7 to 30 VDC max. 1 W
Serial Number - Yes
Response Time - t98% < 30 s at 25 °C
Certificate - Yes, with parameter settings and materials used
ATEX Approval - No
Autoclavable - No
CIP - No
Steam Sterilizable - No
Gamma Irradiatable - No
Operating Temperature Range - 0 to 60 °C