Features: Meter meets with the requirement of international GLP standards. Its features up to automatic calibration, automatic temperature compensation, data storage, timing measurement, clock display, RS-232 output, function setting and self-diagnose etc. There are three different measurement modes: stable display mode; time measurement mode; continuous measurement mode. Equips Conductivity/ATC electrode. Its measurement result is fast and precise. Adopt advanced and original conductivity measurement technology, use 1 pc of conductivity electrode (K=1), it can meet measuring accuracy requirement for range within 0.5μS/cm ~ 200mS/cm by using only one-point calibration. Ability to switch among conductivity, TDS, salinity and resistivity. Multinomial calculation for TDS and salinity to ensure the conversion precision of the full scale. Meter has the only product serial number. Dustproof and waterproof meter meets IP54 standards. All sockets are seal protected by the silica gel caps.
Conductivity
Measuring range
Conductivity:
(0.00~20.00) μS/cm; (20.0~200.0) μS/cm;
(200~2000)μS/cm; (2.00~20.00)mS/cm; (20.0~200.0) mS/cm
Resistivity: (0 ~ 100) MΩ·cm
TDS: (0 ~ 100) g/L
Salinity: (0 ~ 100) ppt
Resolution
0.01/0.1/1μS/cm 0.01/0.1/1 mS/cm
Accuracy
Meter: ± 0.5% FS,connect meter with electrode: ± 0.80% FS
Temp. compensation range
(0 ~ 50)℃ (automatic )
Electrode constant
0.1 / 0.5 / 1 / 5 / 10 / 50 / 100 cm-1
Reference temperature
25℃,20℃ and 18℃
Temperature
Measuring range
-10℃~ 110℃
Resolution
0.1℃
Accuracy
5~ 60℃ range: ± 0.4℃ other range: ± 0.8℃
Other Technical Parameters
Data storage
300 groups