High resolution satisfying European Pharmacopoeia (spectral bandpass: 1.5 nm).
Trace-amount measurement in biotechnological field, etc. can be carried out using optional 50, 25 and 5 micro cells.
Features
Stable optics due to double beam
In double beam design, the energy of the light source is divided into two with a half mirror so that one passes through the reference side, and the other through the sample side, which is unavailable with the single beam design.
Since the reference-side energy is also incident on a detector, photometry is carried out on the basis of this signal. Therefore, an energy change in the light source can be compensated to ensure stable measurement for a long time.
Incorporation of stigmatic concave diffraction grating
The optics of this instrument adopts the Seya-Namioka monochromator widespread as a representative concavediffraction grating monochromator.
Because a concave diffraction grating has both beam condensing and dispersing functions, an optical system canbe configured with fewer mirrors.
In a spectrophotometer, use of fewer mirrors signifies a shorter optical path,thus giving rise to an aberration-free bright optics.
For elimination of the aberrations which were essentially unavoidable in the past, a stigmatic concave diffractiongrating has been developed by applying Hitachi’s original technology. As a result, a higher resolution has been realized.
A higher resolution has been achieved by eliminating coma from the Seya- Namioka monochromator which is the most popular concave diffraction grating monochromator. Its grooving is supported by the only ruling engine in Japan.