Semi-Auto Chemistry Analyzer, ideal for low volume of sample laboratories
Assay Type: End-Point, Kinetic, Fixed-Time, Absorbance, Monochromatic, Bichromatic
Measuring Principles: Colorimetric, Kinetic
Sample Type: Plasma, Serum, Whole Blood
Programming: 80 User defined profiles
Reagent/Sample Handling
Carry-Over: 1%
Incubator: 37°C (programmable), 10 Round Wells
Optical System
Light Source: Halogen Lamp 6V 10W, 2000 Hours
Wavelength: 340, 405, 492, 510, 546, 578, 630nm (±2nm)
Absorption Rate: 0.0 ~ 3.5Abs
Resolution: 0.0001Abs (Displayed), 0.00001Abs (Calculated)
Precision: < 1.0% Flow Cell 32ul, Quartz Glass, Optical Length 10mm
Operating Temperature: 37°C
Temperature Fluctuation: ± 0.1°C
Cuvette: Optional
Control and Calibration
Calibration Mode: Factor, One-Point Linear, Multi-Point Linear
Quality Control: Levy-Jennings
Software
Standby: 30, 60, 90, 120 min (programmable)
Warmup Time: 10min
Display: 7" LCD Color, Touch, 320 x 640
LIS: Mono-Directional (Output Only)
Languages: English, Spanish, Russian
Data Storage: 100,000 (Patients and Quality Control)
Printing: Internal Printer
Working Condition
Power Supply: 100 ~ 240 VAC, 50/60Hz, 120VA
Temperature: 10 ~ 30°C
Humidity: ≤ 85%