Energy accuracy <±1%
Standard Adult/Pediatric Paddles Reusable external adult paddles (pediatric paddles integrated)
Power requirements.
(1) AC Power supply 100V~240V,50Hz/60 Hz
(2) Vehicle Voltage(Option) DC 12V
Internal Battery Backup
Charging time Time-Minimum of 4 hours for full charging
Battery charging During AC Power Supply operation of the device.
Charging /Low Battery Indication Red LED
Battery Capacity 100% of the battery will provide 120-min monitoring; and 35 discharges at 360 joules.
Display
LCD Dimensions 7-inch
Type High resolution true-color graphic
Resolution 480X234
ECG
ECG lead:Ⅰ, Ⅱ, Ⅲ, AVR, AVL, AVF, V1~V6
Protection: With stand 4000vAc/50Hz voltage in isolation and work against electrosurgical knife and defibrillation
Respiration
Measurement method: Thoracic impedance
NIBP
Working mode: Manual/Automatic/stat(5 minutes)
Method Automatic oscillation
Cuff type standard Adult; optional Pediatric/Neonatal
Temperature
Measurement range 20~45℃
Resolution 0.1℃
Pulse Oximetry(spO2)
Display spo2 value, pulse histogram, waveform, Pulse
Range 0~99%for adult, pediatric and neonatal patients
Probe standard adult finger clip
Pulse range
Pulse range 0-255bpm
Standard parameter & configuration
ECG/ RESP/SPO2/TEMP/PR/NIBP/DEFIBRILLATOR