1.1 Gantry system
1.1.1 angle: ±30 degrees can be remote control operation into / out, up / down table and gantry title
1.1.2 Slip ring type: contactless high speed data link slip ring
1.1.3 Detector type: GOS solid detector
1.1.4 Detector row: 32
2.1 X-ray system
2.1.1 Tube anode heat capacity: 5.3MHU
2.1.2 Maximum anode heat dissipation rate of tube: 815 kHU/min
2.1.3 Small focus spot: 0.5 x 1.0 mm
2.1.4 Large focus spot: 1.0x 1.0 mm
2.1.5 Focus to the scanning field and other central distance: 570mm
2.1.6 Minimum tube voltage: 80kV
2.1.7 Maximum tube voltage: 140kV
2.1.8 High voltage generator power
2.1.9 Minimum current: 10mA
2.1.10 Maximum current: 420mA
2.2 Patient Table
2.2.1 Maximum scan range: 1850mm
2.2.2 The highest vertical movement of the table is 990mm, the lowest vertical movement of the table is: 425mm
2.2.3 Maximum table longitudinal speed: 150 mm/s; minimum longitudinal speed: 5 mm/s
2.2.4 The table longitudinal movement accuracy: ±0.25 mm
2.2.5 The maximum patient table load: 250kg
2.3 Console and computer system
2.3.1 The main console
2.3.1.1 Master computer: Lenovo P500
2.3.1.2 Computer memory: 32GB
2.3.1.3 Host hard drive: 5TB
2.3.1.4 DVD-ROM disc recorder system (DICOM compatibility) DVD+R Rom
2.3.1.5 Color LCD flat-panel monitors flicker 1 unit (progressive)
2.3.1.6 Standard DICOM 3.0 interface: standard configuration, with the following functions:
Send / receive Query / retrieval Basic print function Storage
Network interface (RIS / HIS)
2.3.1.7 Laser camera interface
2.3.1.8 Remote diagnosis system and remote diagnosis platform
2.3.1.9 Virtual photo type setting function