1.1 Gantry
1.1.1angle: ±30 degrees, can be remote control operation into / out, rise / fall table and gantry tilt
1.1.2 Slip ring type: contactless high speed data link slip ring
1.1.3 Detector type: GOS solid detector
1.1.4 Detector row: 32
2.1 X-ray system
2.1.1 Tube anode heat capacity: 8.0MHU
2.1.2 Maximum anode heat dissipation rate of tube: 931 KHU/min
2.1.3 Small focus: 0.6 x 1.2 mm
2.1.4 Large focus: 1.1 x 1.2 mm
2.1.5 Focus to the scanning field and other central distance: 570mm
2.1.6 Minimum tube voltage: 80kV
2.1.7 Maximum tube voltage: 140kV
2.1.8 High voltage generator power
2.1.9 Minimum current: 10mA
2.1.10 Maximum current: 667mA
2.2 Patient Table
2.2.1 Maximum scan range: 1850mm
2.2. The highest vertical movement of the table is 990mm, the lowest vertical movement of the table is 425mm
2.2.3 Maximum table longitudinal speed: 150 mm/s; minimum longitudinal speed: 5 mm/s
2.2.4 The table longitudinal movement accuracy: ±0.25 mm
2.2.5 The maximum patient table load: 250kg
2.3 Console and computer system
2.3.1 The main console
2.3.1.1 Master computer: HP Z820
2.3.1.2 Computer memory: 32GB
2.3.1.3 Host hard drive: 7TB
2.3.1.4 CD-ROM disc recorder system (DICOM compatibility)
2.3.1.5 Color LCD flat-panel monitors flicker 1 unit (progressive)
2.3.1.6 Dtandard DICOM3.0 interface: standard configuration, with the following functions:
Send / receive Query / retrieval Basic print function Storage
Network interface (RIS / HIS)
2.3.1.7 Laser camera interface
2.3.1.8 Remote diagnosis system and remote diagnosis platform
2.3.1.9 Virtual photo type setting function
2.3.1.10 Console color 3D display
2.3.2 Advanced workstation