High accuracy and speed for daily machine parameter verification
The finest resolution for spot characterization and Machine QA
Daily machine parameter verification used from commissioning to machine QA, without compromise between high accuracy and speed
-Optimized for pencil beam scanning
-High resolution scintillator-based sensor
-Active surface of 30x30 cm² and effective resolution of 0.5 mm
-Single shot and movie mode measurements
-Compatible with myQA
-Dicom RT export supported
-IBA Dosimetry Proton Lynx Product
-Sphinx with Lynx Publikation
-Placidi et.al. “Range Resolution and Reproducibility of a Dedicated Phantom for Proton PBS -Daily Quality Assurance.”
-Zeitschrift Für Medizinische Physik 28, no. 4 (December 1, 2018): 310–17.
The Lynx detector consists of a box composed by a scintillating screen of gadolinium-based plastic material which converts the energy lost by ionizing radiation into photons. This light is reflected to photodiodes of a CCD camera outside the irradiation field. To avoid the saturation of the camera, iris aperture can be changed, as well as the gain. The detector signal from the CCD camera is digitalized at a 10-bit depth.
The Lynx detector is easy to setup. It can be operated on the treatment coach. Data is transferred to a PC via Ethernet (previous version used Firewire).
Lynx is supported by the following software solutions:
-Lynx2D
-myQA Lynx plug-in