The Model 2024PRO Series is a 300-1600ml microprocessor-controlled unit providing the latest in electronic updates. Developed as a compliment to your already owned 2002 PRO, all parts, accessories, and consumables for the Model 2024PRO and Model 2002PRO are completely interchangeable.
All PRO series ventilator purchases now include a Ventilator Accessory Kit.
The single valve turns a total of nine and a half times; the first two and a half turns utilize the slight, minute changes for the pediatric animal patient while the remaining rotations are to be used with the adult and extra-large sized bellows and housings.
Low 50 psi Supply Gas Alarm – The Low Supply Gas Alarm is activated when the supply gas pressure drops below 35 psi (2.4 bar). During an alarm condition, the red light on the front panel is illuminated and a steady, continuous tone is heard. This alarm automatically resets when the pressure increases above 40 psi (2.7 bar).
All Hallowell Ventilators are equipped with dual airway pressure alarms that monitor and sound when the Peak Inspiratory Pressure (PIP) of any breath either exceeds the preset Maximum Working Pressure Limit (MWPL) or fails to reach 6 cm H2O as in the case of a disconnect.
Volume – Cycled microprocessor-controlled ventilator.
Time cycled – Ensures consistently spaced breaths. The possibility of stacking breaths or auto-PEEP is eliminated by keeping the I:E ratio constant at 1:2. This means that regardless of the respiratory rate, setting two-thirds of the breath will remain for expiration. The respiratory rate is adjustable from 6 – 40 breaths per minute (BPM).