Stainless steel funnel SS316L, stainless steel filter head SS316L
Glass funnel, glass filter head , PTFE sieve
Stainless steel holder (SS316L)
Stainless steel valve
Stainless steel lid SS316L
Clamp of anodized aluminum
Membrane Selection. The selection of the membrane is the key to achieve the best filtering effect. The membrane material must be suitable for the different characteristics of the filter medium and has to be chemical resistant against it. Available with pore size : 0.22μm , 0.45μm , 0.8μm and more . The diameter of the membrane is Φ47mm or Φ50mm .
All the manifolds can be autoclaved at 121°C for 20-30 minutes or sterilized with epoxyethane etc.
operation
Attach the glass filter head on the SS holder (all SS filter manifolds have head and holder connected in one piece) and turn carefully to fit with the holder.
Put the membrane filter on the sieve , attach the funnel to the filter head , and attach the clamp .
Connect the rubber tube to the vacuum nozzle of pump , the other side to the collecting bottle .
Connect the rubber tube to the nozzle of the manifolds vacuum filtration , the other side to the nozzle of thecollecting bottle.
Fill the solution into the funnel, open the valve and start up the vacuum pump . The solution flows through the membrane in the filter head through the holder into the collecting bottle due to the action of vacuum .