6 channel ECG machine with interpretation
5.0 inch color TFT LCD display
Support U disk storage
Manual and Auto working modes
Input impedance: ≥50MΩ
Skin voltage tolerance: ≥±500mV
Description
The 6 channel ECG machine, using digital isolation technology and digital signal processing, has two working mode, manual and auto working modes. In addition to the baseline stabilization inspection and full alphanumeric silicon keyboard, it also supports U disk storage.
6 channel ECG machine with interpretation
5.0 inch color TFT LCD display
Simultaneous 12 leads acquisition and 1, 1+1, 3 and 6 channel (Manual/auto) recording with high resolution thermal printer
Manual/auto working modes
Use digital isolation technology and digital signal processing
Baseline stabilization inspection
Full alphanumeric silicon keyboard
Support U disk storage
Specifications
Lead - Standard 12 leads
A/D converter - 14bit
Working mode - Manual/auto
CMMR - ≥100dB (with AC filter)
Input circuit current - ≤0.1μA
Patient current leakage - <10μA
Skin voltage tolerance - ≥±500mV
Frequency response - 0.05 Hz~250Hz
Sensitivity threshold - ≤20μV
Page speed - 6.25, 12.5, 25, 50mm/s (±3%)
CPU - 32 bits
Electricity - AC 220V (±10%), 50Hz±1Hz, Power ≤60VA,
AC 85~265V, 50/60Hz
DC14.8V(2200mAh), rechargeable Li-ion battery
Acquisition - Simultaneous 12 leads
Sampling rate - 1000Hz
Interference filter - AC filter and EMG filter
Input circuit - Floating, protection circuit against defibrillator effect
Input impedance - ≥50MΩ
Calibrating voltage - 1mV±2%
Time constant - ≥3.2s
Sound emission level - <15μVp-p
Signal channel interference - ≤0.5mm
Paper - 112mm*20m, roll