AG-IRW002 Medical Infant Radiant Warmer With Three Control Modes Infant warmer on casters
Function features
1. Microprocessor-based servo controlled temperature system
2. The microprocessor-controlled unit allows for three modes: pre-warm, manual and automatic
3. APGAR timer to record therapy time
4. Audible and visual Alarm functions :
Power failure, temperature deviation, temperature sensor failure, skin temperature failure, over temperature
5. LED observation light adjustable, brightness adjustable
6. Heater head can be adjusted ±90°horizontally
7. Convenient X-ray cassette board under infant bed
8. Tilt angle of infant bed is adjustable
9. Four Transparent protectors folded with 2 infusion seals
10. With RS-232 connector,2 trays, and 2 drawers
Technical data
Power supply:AC110/220V 60/50Hz
Power input:1000VA
Operating conditions - Environment temperature:18℃~30℃
Environment relative humidity:30%~75%
Atmospheric Pressure:700~1060hpa
Environment air velocity of flow:<0.3m/s
Skin Temperature control range:32℃~38℃
Skin temperature control accuracy:≤0.5℃
Skin temperature sensor accuracy:±0.3℃
Mattress temperature uniformity:<2℃
Infant Bed tilt angle:±10°
Warm-up time(from 25℃):<30min
Heater head horizontal:±90°
Distance from heater to mattress:80cm
Mattress size:67cmX54
Timer range:0~99h59min
Package
Each unit is packed with one plywood case
Size:118*90*87cm
N.W.:75kg
G.W.:100kg
Transport and storage
Environment temperature:-40℃~+55℃
Environment relative humidity:≤95%
Atmospheric Pressure:50-1060hpa