AG-IIR001B Medical NICU Obstetrics Infant Incubator With Cupboard For Baby
Infant Incubator On Casters
Specification:
Model:AG-IIR001B
Power:~ 220V 50Hz
Input power:650VA (including power output socket power: 200VA)
Temperature:20 ° C ~ 30 ° C
Humidity:30% RH ~ 75% RH
Air pressure:700hPa ~ 1060hPa
Air velocity:less than 0.3m/s
Box temperature:25 ° C ~ 37 ° C; 37.1 ° C ~ 38 ° C
Warm-up time:≤ 45 minutes
Box temperature over temperature:38 ° C (power off the heater) sound and light alarm
The second thermal cut-off device over-temperature alarm:40 ° C (cut off the heater power supply) sound and light alarm
Box temperature deviation:±3.0 °C (power off the heater at +3.0 °C) sound and light alarm
Sensor alarm:When the box temperature sensor is open or short circuit, the heater power supply is cut off, and the sound and light alarm is given
Power interruption alarm:power interruption, sound and light alarm
Fan failure:The fan stops and its sensor system fails, the sound and light alarms, and the heater power is cut off
The maximum carbon dioxide concentration:0.4% (tested in accordance with the provisions of Article 105.1 of GB11243-2008)
The noise level of the whole machine (including phototherapy device) in the baby cabin(During normal operation):Less than 55dB (A weighted)