This kit is used for the combined qualitative detection of SARS-CoV-2, influenza A virus, influenza B virus, adenovirus, mycoplasma pneumoniae, chlamydia pneumoniae, respiratory syncytial virus and parainfluenza virus(Ⅰ, II, III, IV) in throat swabs and sputum samples, human metapneumovirus, haemophilus influenzae, streptococcus pneumoniae, klebsiella pneumoniae, staphylococcus aureus, pseudomonas aeruginosa, legionella pneumophila and acinetobacter baumannii.
Technical Parameters
Storage
≤-18℃ In dark
Shelf-life
12 months
Specimen Type
Oropharyngeal swab samples, Sputum swab samples
CV
≤5.0%
Ct
≤40
LoD
300 Copies/mL
Specificity
the cross-reactivity study shows that there is no cross-reactivity between this kit and rhinovirus A, B, C, enterovirus A, B, C, D, human metapneumovirus, epstein-barr virus, measles virus, human cytomegalovirus, rotavirus, norovirus, mumps virus, varicella-band herpes zoster virus, bordetella pertussis, streptococcus pyogenes, mycobacterium tuberculosis, aspergillus fumigatus, candida albicans, candida glabrata, pneumocystis jirovecii, cryptococcus neoformans and human genomic nucleic acid.
Applicable Instruments:
Applied Biosystems 7500 Real-Time PCR System
Applied Biosystems 7500 Fast Real-Time PCR Systems
QuantStudio®5 Real-Time PCR Systems
SLAN-96P Real-Time PCR Systems
LightCycler®480 Real-Time PCR system
LineGene 9600 Plus Real-Time PCR Detection System
MA-6000 Real-Time Quantitative Thermal Cycler
BioRad CFX96 Real-Time PCR System
BioRad CFX Opus 96 Real-Time PCR System