Design: the instrument is widely used in medicine and health, clinical testing, biochemistry, environmental protection and other
industries, and is one of the analytical instruments reused in physical and chemical laboratories. Used for quantitative or qualitative
analysis of substances.
Working principle: matter has selective absorption of light, resulting in an absorption effect when exposed to light. Different substances
have different absorption spectra, and when monochromatic light passes through a solution, its energy will be weakened
due to absorption, and the degree of energy attenuation is proportional to the concentration of the substance.
This series of instruments is based on the principle of colorimetry to qualitatively and quantitatively analyze samples, and within a
certain concentration range, it conforms to the Lambert-Beer law.
Technical parameters
▪▪ Optical system Achromatic Czerny-Turner (1200L/mm grating)
▪▪ Wavelength range 195-1020nm
▪▪ Wavelength accuracy ±2nm
▪▪ Spectral bandwidth 5nm
▪▪ Wavelength repeatability 1nm
▪▪ Stray light ≤0.3%T
▪▪ Photometric accuracy 0.5%T
▪▪ Photometric repeatability 0.2%T
▪▪ Photometric range 0-200%T,-0.301-2.000A
▪▪ Drift ≤0.2%T
▪▪ Baseline linearity ±0.004A/h @500nm
▪▪ Noise 100%≤0.3%T
▪▪ Stability ±0.004A/h @500nm
▪▪ Way of working T,A,C
▪▪ Display range 0-200%T, -0.3-2A
▪▪ Display system 4-digit LCD LED display
▪▪ Wavelength drive manual
▪▪ Printer (interface) Optional thermal printer (parallel port)
▪▪ Energy (light source) Imported tungsten lamp, imported deuterium lamp
▪▪ Light switching method manual