Product application positioning: Hospitals, maternity centers, neonatology departments.
Design: The product is designed to provide a treatment environment with clean air and appropriate temperature for low-severity infants,
critically ill infants and premature infants.
Infant incubator control system adopts computer technology to implement servo control of the temperature in the box (box temperature/skin temperature), according to the set temperature and measured temperature proportional heating control. The internal air uses the principle of heat convection to adjust, to create a suitable air, temperature and humidity, warm and comfortable, similar to the mother's womb good environment, so as to cultivate and care for the baby.
TECHNICAL PARAMETERS:
Mattress to thermostat height 450 mm
Front side door panel 826×310 mm
Mattress tray size 630×360 mm
Crib tilt adjustment ±10° stepless
Power supply A C 220V,50Hz
Power requirement ≤400VA
Temperature control mode A ir and skin servo control
Air temperature control range 20.0°C-37.0°C
Temperature overcontrol mode to 37.1°C-39.0°C
Temperature control range 20.0°C-37.0°C
Temperature overcontrol mode to 37.1°C-38.0°C
TECHNICAL CONFIGURATION:
Accuracy of temperature sensor ≤0.3°C
Display range 5%-99%
Temperature fluctuation ≤0.5°C
Mattress temperature uniformity ≤0.8°C Rise time: ≤45 minutes
Thermostat noise level ≤55dB
Ambient temperature +20°C-+30°C
Ambient relative humidity 30%-75%
Ambient wind speed <0.3 m/s
Crib maximum load 5KG
Maximum load-bearing capacity of infusion stand 2KG