Absorption Spectrometry, Turbidimetry
Test Method: End-point, Fixed-time, kinetic
Test Sequence:
Run tests in strict accordance with sample sequence,
Insert emergency at any time
Reaction Time: Max 15minutes
Reaction Volume: Mini 150μl
Cuvette: 81 cuvettes
Optical System
Light Source: Halogen-tungsten lamp, 12V/20W
More than 2000hrs service time
Spectrometric Method: After spectrometric method
Wavelength: 8 wavelengths, selectable among 340-800nm
Detector: Photodiode array
Absorption Range: 0-4.0Abs
Optical System: Integrated full closed optical system
Sample system
Sample Probe: Polishing for both interior and exterior, Liquid level detection, anti-collision protection and inventory checking; probe with special cleaning solution to prevent contamination
Mixing System: Separate mixer, mix the sample and reagent completely
Others
Water Consumption: less than 4L/hour
Power Supply: AC 220 ±22V 50HZ ±1HZ
Power: Max 200VA
Ambient Temperature: 15°C-30°C
Reagent System
Reagent Probe: polishing for both interior and exterior, Liquid level detection, anti-collision protection and inventory checking, have automatic stop and alarm prompt for an obstacle
Reagent Bottle:
Bottles with 15ml and 30ml are available,
dead volume ﹤0.5ml
Temperature Control System
Heating Mode:
Solid constant temperature closed direct heating
Heating Time: less than 3minutes
Reaction Temperature: 37±0.2°C
Temperature Fluctuation: ± 0.1°C
Calibration and QC
Calibration Mode:
Linear, Logit-4p, Logit-5p, Spline, Factor, Polynomial
Automatically check calibration curve and fitting the best calibration curve
QC Type: Multi-type
QC Rules: Multi-rule; default: Westguard