Additional ABP Recorder for an earlier purchased Holter ECG or ABP System. The diagnostic device operates with rechargeable or alkaline batteries, being suitable for 50-hour BP monitoring with up to 300 BP readings, using oscillometric technology.
BP measurement method:Oscillometric with step deflation, Disatolic values correspond to Phase 5 Korotkoff Sounds
BP measurement range:20 – 280 mmHg (max inflation 300 mmHg)
Pressure Transducer Accuracy:± 3 mmHg
Pulse Rate Accuracy: ± 2% or ±3 BPM, whichever is greater
Calibration: The device should be verified on a yearly interval
Safety Systems: Max. inflation pressure is limited to 300 mmHg;
Auto safety release valve for power failure.
Maximum cuff inflation time is limited to 75 seconds.
Duration of blood pressure reading is limited to 130 seconds.
Data Connector: Mini USB B type connector
Min. Power supply2 x 1.2 V AA rechargeable batteries of minimum
2500 mAh capacity or 2 x 1.5V AA high quality
alkaline batteries
Size: - Approx. 125 x 70 x 33 mm
Weight: Approx. 250g (incl. Batteries)
Screen: No screen
Clinical Accuracy:Meets accuracy requirements of ANSI/AAMI
SP10:2002(R)2008, EN1060-4:2004 and ISO 81060-2:2009.
Patient button:Yes
Storage Card:built-in 2GB (uSD), more than 4000 measurement
Interval settings: In intervals of 10 to 240 minutes (adjustable every minute)
Recording Period (hours): 8-336 hours with battery replacement (~ 48 hours recording with a set (2) of batteries)