Featuring LANDAUER’S leading Optically Stimulated Luminescence (OSL) technology, the InLight Environmental radiation dosimeter requires no preparation and provides complete monitoring and reanalysis capabilities for dose verification and archiving. It meets the standards of the American National Standards Institute and Health Physics Society and its case is sealed in a waterproof plastic pouch.
Advanced Design
InLight dosimeters are built on an assembly of a case component with metal and plastic filters along with a four-positioned aluminum oxide detector slide component.
Both the case and slide are uniquely bar coded with serial numbers for chain of custody and sensitivity identification. InLight dosimeters offer reanalysis capabilities, precision with a wide dynamic range of measurement, and long-term stability. The InLight Basic dosimeter consists of the case and slide for use with a clear plastic holder.
The enhanced LANDAUER holder is designed to accommodate the optional CR-39 for neutron detection, the optional imaging component, client defined labels and the case assembly and slide. The case component has an open window with aluminum, copper and plastic filters. The imaging component renders unique filter patterns to provide qualitative information about conditions during exposure. Dosimeter labels can be vertical or horizontal and offer numerous graphic and text fields definable by the client to meet the administrative needs of a radiation monitoring program.
The environmental dosimeter is designed to meet ANSI N545 Standard and ANSI/HPS
Standard N13.37.