The multigas calibrator with GPT, ozone and self re-generable zero air generator, Sonimix 3012/3022, is an instrument designed to calibrate manually or by remote control up to 6 monitors installed in an air pollution monitoring station, such as SO2, HC, BTX, CO, NOx and O3.
The Sonimix 3012/3022 is equipped with
– an inlet for high concentration mixtures of diluted gases such as NO, NO2, SO2, CO, HC, BTX or mixtures of several compatible polluants in the same cylinder.
– a zero air generator composed of an internal compressor with self re-generable filters that removes moisture, O3, NOx, SO2, HC, BTX and optionally CO and CH4.
– a binary gas dilutor with two inlets (diluted gas and air), based on sonic nozzles (ISO 6145-6 standard)
The ozone generator is regulated by UV-light feedback and compensated from the atmospheric pressure and ambient temperature variations, with alarm on the drift of UV-light as well as the circuits for the GPT.
-Accuracy – (0,6% rel.)
-Best quality price ratio
-No maintenance nor pressurised Air cylinder with its included self re-generable zero air -generator
-Sonic nozzle technology
-Remote control
-Ergonomic with LCD display
-Air pollution monitoring
Dilution point -
3 - 10
Dilution range -
1/166, 1/250, 1/500
(100, 200, 300 ppb with 50 ppm at inlet)
1/200 to 1/2000 or 1/100 to 1/1000
Dilution principle -
Sonic nozzle
Dilution gas line flow accuracy -
0,6% relative
Repeatability on flow -
Better than 0.3% relative
Warm up time -
< 45 min
Stabilization time -
3 min
Operating temperature -
15°C – 40°C
Dimensions -
19″ 3HE/84 TE , 500mm deep
portable casing in option
Net weight -
20 Kg