The LogTag® UTREL30-16 Dual-Channel Low Temperature Recorder measures and stores up to a total 32,000 temperature readings.
It features a large easy-to-read display and is designed to show ‘at a glance’ if temperature excursions have occurred during the current day and up to the previous 29 days.
The alarm is triggered if temperature readings are outside pre-set alarm limits which is indicated by a large X on the display and, if configured, the audible alarm will sound.
Specifications
Storage Temperature Range
0°C to +40°C (32°F to +104°F)
Rated Temperature Resolution
Better than 0.1°C (0.2°F) for measurements -80°C to 10°C (-112°F to 50°F)
Better than 0.2°C (0.4°F) for measurements 10°C to 20°C (50°F to 68°F) and -90°C to -80°C (-130°F to -112°F)
Better than 0.3°C (0.5°F) for measurements 20°C to 30°C (68°F to 86°F)
Better than 0.5°C (0.9°F) for measurements 30°C to 40°C (86°F to 104°F)
LogTag Analyzer currently displays to one decimal place of °C or °F. The native resolution is what is stored in the LogTag®.
Sensor Reaction Time
Typically less than 2 minutes(T90) in moving air (1m/s) for ST10S type
Recording Capacity
32,000 total log memory. Provides 16,000 temperature readings per channel
Day summary statistics memory (for display on LCD): up to 30 days of Min/Max/duration and alarm trigger statistics
Sampling Interval
Configurable from 30 seconds to 18 hours. (Typically 5 minute logging)
Alarms
Up to 6 individually configurable alarms for each channel
Possible to set alarms separately for Channel 1 and Channel 2
Logging Start Options
Push button start (with configurable start delay from 1 minute to 72 hours) or specific date & time