This direct reading dosimeter is a rugged instrument which measures accumulated quantities of gamma and X-ray radiation up to 5 R. Applications include personal and environmental monitoring. The low energy feature has hospital applications, including fluoroscopy, portable radiography, and angiography. This pocket size instrument is lightweight, and has a sturdy clip to attach to an individual's pocket.
To ensure accurate readings, the AT-725 must be zeroed before use with a dosimeter charger, such as the AT-909 or The Charger
Features
0 to 5 R
Lightweight
Hermetically Sealed
Sturdy Pocket Clip
Meets ANSI N13.5 & N322-1997
Specifications
Detector - Fiber electrometer mounted in an electrically conductive plastic ion chamber
Detector Housing - Very low permeability plastic, hermetically sealed
Accuracy - Within 10% of true exposure
Rate Response - Dose rate independent for gamma and X-ray radiation
Size (D x L) - 1.5 x 12.4 cm (0.6 x 4.5 in.)
Weight - 25 g (1.0 oz)
Warranty - 2 year limited warranty
Qualifications - Meets ANSI Specifications N13.5 and N322-1997