Replacement For
Analytical Industries [PSR-11-915-1]
Drager [6803290, 6850645, Narkomed, Fabius GS, Tiro]
Hewlett Packard [15201A]
Invivo [Magnitude]
Mercury Medical [10-103-12]
Pacifitech [PT-24]
Teledyne [R31]
VTI [101372, 101372]
Specifications
Range of Measurement: 0 to 100% O2
Output Range: 12.2 to 18.1 mV
Zero Offset: ≤ 0.50 mV
Response Time: ≤ 15 seconds at 23±2°C
Linearity: ± 1.0%
Expected Lifetime: > 2,000,000 %O2 hours
Operational Temperature Range: 5° to 40°C
Storage Temperature Range: -15° to 50°C
Humidity Range: 5% to 95% RH, non-condensing
Interference: Per EN ISO 80601-2-55