B-Scan
Scanning method: mechanical sector scan.
Gain control: 98dB.
TGC: -30dB – 0 dB dynamic, manually adjusted (5 scopes).
Display mode: B, B+B, B+A, A.
Scanning scope: depth 34mm-60mm, scanning angle: 53º.
Dynamic review: 5.6sec/56 images, single or multiple.
Gray scale: 256.
Pre-image processing: frame even.
Post-image processing: 4 groups of curves (liner, logarithmic, exponential, S).
Color Display: 8 color modes.
Electronic caliper: electronic cursor for multiple distance measurement,
Accuracy no less than 0.25mm.
Area measurement: Accuracy no lower than ± 0.0625mm2.
Delayed depth control: 0mm~15mm.
Built-in Calendar & Clock.
A-Scan
Resolution: 0.01mm
Biometry Accuracy: ≤±0.06mm
Measuring range (AL): 15-39mm.
Total gain: 98dB, user adjustable gain range: 0-55dB.
Measuring parameter: anterior chamber depth, lens thickness, vitreous length and axial length.
Measuring mode: 5 (normal, aphakic, special, dense cataract and manual measuring).
Measuring method: immersion, contact.
IOL calculation: SRK-T, SRK-II, BINK-II, HOLLADAY, HOFFER-Q, HAIGIS. Any two calculations can be compared and displayed on the same screen.
Calculation of 8 groups’ average with standard deviation (S.D.).
50 groups of results saved permanently.
groups of IOL constants can be saved.