It is mainly used on delivery room, baby wards, NICU that the baby’s maximum weight up to 10Kg.
Inside an alarm system of differential pressure. When alarming, the safety valve starts automatically and produces the mixture gas continuously to ensure the safety of patients.
Adjust the Oxygen concentration from 21% to 100% optionally.
The range of flow adjustment is 5LPM~15LPM.
The Max-P and PIP are safe, controlled and accurate.
The PEEP is constant and accurate.
Simple operation and flexible application.
The output pressure would not be influenced by the operator experience, training, attention and fatigue.
Technical Parameter
Pressure gauge :- 2˜10kPa (-20˜100cmH20).
Max-P : @0˜15LPM 6kPa (60cmH20).
PIP : @5˜15LPM 0.2˜5.7kPa (2˜57cmH20).
PEEP : @5˜15LPM 0.03˜2.3kPa (0.3˜23cmH20).
RESP-17 : come with inbuilt Air / Oxygen blender.