horizontal blood bank refrigerator
Cabinet Construction
Bacteria resistant stainless interior, exterior, and door handle(s)
Ergonomic door handle is padded for comfort
Superior engineered frame for maximum rigidity
Dual-pane glass door(s) with ADA compliant full-length handle
Sure Seal self closing door with magnetic closure
LED lighting with control panel mounted switch
Auto ON/OFF1
Minimum of 4” (100mm) non-CFC foamed urethane insulation
Swivel locking casters
Access port located in top of cabinet for external monitoring probe(s)
Recessed floor contains spills for easy clean-up
Hospital grade plug and cord
Storage
Stainless steel pull-out drawers
Durable, liquid-tight construction for containment of spills
Stainless steel drawer standards guard against rust and corrosion
Easy access to inventory with fully extendable drawer slides
Scratch Guard edge protectors keep glass doors from being scratched by drawers
Tube Guard drawer shields minimize the risk of tube damage
Chart Recorder
6” (150 mm) chart recorder with 7-day circular charts
Inkless, pressure-sensitive chart paper (-5°C to +20°C)
Battery backup for 48 Hours
Button controls for chart changes and calibration
Refrigeration System
Heavy-duty, air-cooled refrigeration system with non-CFC R134A refrigerant provides high efficiency and is environmentally friendly.
Forced-air circulation maintains chamber uniformity of ±1°C and provides quick recovery after door openings
No defrost cycle required to maintain constant temperature
Auto condensate evaporation
Evaporator fan(s) shut off during door openings to maintain stable temperatures