They are units that provide all the infrastructure services needed in intensive care departments of hospitals, such as “medical gases, electricity and weak current (data communication)”, necessary services and device connections to the patient in the most comfortable and fastest way.
Intended Use
• Placement and spatial positioning of medical devices and accessories used for diagnosis and therapies.
• Supply of medical devices powered by electrical energy, medical gases and vacuum.
Environments of Use
The product is intended for medical environments, in particular emergency rooms, intensive care units and post-operative care units.
Do not use the device in the following environments
• Areas where the oxygen concentration is above 25 percent by volume, where flammable gas mixtures or explosive gas mixtures may occur.
• Places where magnetic field applications are performed (e.g. magneticresonance tomography)
Standards
✓ EN 62366-1 : 2015 ✓ EN ISO 15223-1 : 2021 ✓ EN ISO 20417 : 2021
✓ EN ISO 11197 : 2019 ✓ EN ISO 14971 : 2020 ✓ EN 5359 : 2012
✓ EN 60601-1 : 2009 ✓ EN 60601-1-2 : 2016 ✓ HTM 02-01
High Ceiling Connection Carcass System
Applied for ceiling connections over 3 meters. Provides excellent strength and durability. Resistant against oxidation and corrosion.
Anaesthesia Device Connection Apparatus
Provides fast and secure connection in cases where the anaesthesia device needs to be upgraded by connecting to the unit.
Ceiling Connection Flange Gas Electrical Connections
During installation, maintenance and service, gas and electricity can be easily connected and disconnected from the unit at any time.