Safer than disinfection, hot-air sterilisation of containers and instruments
Standard delivery Steriliser
The indispensable safety feature for this hot air steriliser: Setpoint Wait. This means that the programme for sterilisation only starts when the set temperature has been reached. This feature can also be used with freely positionable Pt100 temperature sensors. Here the sterilisation time only begins when the set temperature has been reached at all measurement points, and reliable sterilisation is guaranteed at all times.
Control technology
ControlCOCKPIT: TwinDISPLAY. Adaptive multifunctional digital PID-microprocessor controller with 2 high-definition TFT-colour displays.
Language setting: German, English, Spanish, French, Polish, Czech, Hungarian
Timer: Digital backwards counter with target time setting, adjustable from 1 minute to 99 days
Function HeatBALANCE: adapting the distribution of the heating performance of the upper and lower heating circuit from -50 % to +50 %
Function SetpointWAIT: the process time does not start until the set temperature is reached
Calibration: three freely selectable temperature values
adjustable parameters: temperature (Celsius or Fahrenheit), fan speed, air flap position, programme time, time zones, summertime/wintertime
Ventilation
Fan: forced air circulation by quite air turbine, adjustable in 10 % steps for each segment individually
Fresh air: Admixture of pre-heated fresh air by electronically adjustable air flap
Vent: vent connection with restrictor flap
Communication
Documentation: programme stored in case of power failure