Unparalleled homogeneity of temperature and humidity. For stability testing according to ICH - WHO - EMA - ASEAN - GMP - GLP – GCCP.
Standard delivery Climate chamber
On this page, you can find all the essential technical data on the Memmert climate chamber ICH. Our customer relations team will be pleased to help if you want further information.
Temperature
Working temperature range: (not suitable for long-term storing at sub-zero temperatures. During permanent operation, the glass door may ice over)
Setting accuracy temperature: 0.1 °C
Temperature sensor: 2 Pt100 sensors DIN Class A in 4-wire-circuit for mutual monitoring, taking over functions in case of an error
Humidity
Humidity: humidity supply with distilled water from external tank by self-priming pump
Humidification: humidification by hot steam generator
Humidity: active humidifying and de-humidifying adjustable from 10-80 % rh with digital display of relative humidity - resolution of display 0.1 %, setting accuracy 0.5 %
Control technology
ControlCOCKPIT: TwinDISPLAY. Adaptive multifunctional digital PID-microprocessor controller with 2 high-definition TFT-colour displays.
adjustable parameters: temperature (Celsius or Fahrenheit), fan speed, relative humidity, programme time, time zones, summertime/wintertime
Language setting: German, English, Spanish, French, Polish, Czech, Hungarian
Timer: Digital backwards counter with target time setting, adjustable from 1 minute to 99 days