Unparalleled homogeneity of temperature and humidity. For stability testing according to ICH - WHO - EMA - ASEAN - GMP - GLP – GCCP.
Standard delivery Climate chamber
On this page, you can find all the essential technical data on the Memmert climate chamber ICH. Our customer relations team will be pleased to help if you want further information.
Temperature
Setting accuracy temperature: 0.1 °C
Temperature sensor: 2 Pt100 sensors DIN Class A in 4-wire-circuit for mutual monitoring, taking over functions in case of an error
Humidity
Humidity: humidity supply with distilled water from external tank by self-priming pump
Humidification: humidification by hot steam generator
Humidity: active humidifying and de-humidifying adjustable from 10-80 % rh with digital display of relative humidity - resolution of display 0.1 %, setting accuracy 0.5 %
Control technology
ControlCOCKPIT: TwinDISPLAY. Adaptive multifunctional digital PID-microprocessor controller with 2 high-definition TFT-colour displays.
Language setting: German, English, Spanish, French, Polish, Czech, Hungarian
Timer: Digital backwards counter with target time setting, adjustable from 1 minute to 99 days
Function SetpointWAIT: the process time does not start until the set temperature is reached
Calibration: three freely selectable values each, temperature/humidity/CO2
adjustable parameters: temperature (Celsius or Fahrenheit), fan speed, relative humidity, programme time, time zones, summertime/wintertime