CO2 Incubator ICO150med Safety at all times: High-end functions for the protection of cell cultures, bacteria cultures or tissue cultures.
CO2 Incubator ICO150med
Standard delivery CO2 Incubator
On this page, you can find all the essential technical data on the Memmert CO2 incubator ICOmed. Our customer relations team will be pleased to help if you want further information. If you should require a customised special solution, please contact our technical specialists at myAtmoSAFE@memmert.com.
Temperature
Working-temperature range5 °C above ambient temperature up to +50 °C
Standard sterilisation programme: 60 minutes at 180°C (without removing the sensors)
Setting accuracy temperature0.1 °C
Temperature sensor2 Pt100 sensors DIN Class A in 4-wire-circuit for mutual monitoring, taking over functions in case of an error
Temperature variation in chamberat + 37 °C (to DIN 12880:2007-05) +/- 0.3 K
Temperature fluctuations with timeat 37 °C (to DIN 12880:2007-05) +/- 0.1 K
Humidity
Humidity control (standard)Humidity limitation thanks to a Peltier element; when water dish is full and inserted, the Peltier element limits the value of relative humidity in the interior to 93 % rh +/- 2.5 %
Setting accuracy humidity0.5 % rh
Setting range active humidity control (with option K7)40 to 97 % rh and rh-Off