Maximum accuracy and no risk of error from incorrect coding
HIGH-CONTRAST DISPLAY
Large digits and backlit screen for a better readability of glucose results
MARKERS
4 markers for before and after meals, exercise and general purpose
NFC download is easy and fast!
NFC (Near Field Communication) enables the download of blood glucose and ketone results from your meter to your smartphone or tablet, simply by holding them back-to-back for a few seconds.
Follow these simple steps to setup and begin NFC transmission.
SMART DESIGN
Modern looks and large, handy strips for an enhanced user experience while testing
Test strip
Sample
Fresh capillary whole blood
Sample volume
0.5 µL
Measuring principle
Electrochemical
Enzyme system
Glucose oxidase
Test range
20-600 mg/dL (1.1-33.3 mmol/L)
Haematocrit range
10-70% (haematocrit compensated)
Testing conditions
Temperature: 5-45 °C (41-113 °F)
Relative humidity: 20-90% (no dew condensation)
Icodextrin interference
The GLUCOFIX TECH system is suitable for use by
people on peritoneal dialysis using solutions containing
icodextrin
Device standard
EN ISO 15197:2015
Batteries
Two 3.0 V lithium batteries (CR2032)
Battery life
Approx. 1000 tests
Data transmission
To PC (via USB cable)
To mobile devices (via NFC or external Bluetooth)
Averages
Over 1, 7, 14, 40, 60, 90 days
Dimensions
85 mm x 56 mm x 18 mm
Alarms
Up to 3 settable acoustic alarms