Our MAS-100 CG Ex® system is for use when there is a need to test compressed gases for microbial contamination. In fact, it is the only compressed gas microbial air sampler approved for use in Zone 2 explosion hazard areas. Using its two sensors, one to measure mass flow and the other for pressure, the unit collects samples under the pressure of the gas supply. The unique MAS-100 CG Ex® system makes compressed gas sampling safe, easy and accurate.
Collects samples under gas pressure; after sampling, gradual decompression occurs, avoiding possible damage to microorganisms
Sampling at two different flow rates: usable for low flow applications
The only compressed gas microbial air sampler approved for use in Zone 2 explosion hazard areas
Factory calibration only: reduces the risk of errors
Unique flow sensor technology: no manual adjustment needed for different gas types after selection of the gas
Automated collection process: saves you time
Pre-programmed for a range of different gases: no further calculations needed
Can store up to 10 gas protocols
Height - 32.5 cm
Length - 37.0 cm
Width - 11.0 cm
Weight (without Sampling Head) - 10 kg
Material Box - Coated aluminum
Nominal Flow Rates - 100 liter/min. ±5.0% (over the pressure range) of 1.5 bar to 10 bar (absolute) 50 liter/min. ±5.0% over the pressure range of 1.5 bar to10 bar (absolute)
Standard Sampling Volumes - 50, 100, 250, 500, 1000 liter
Freely definable sampling volumes - 1 to 2000 liter, volumes individually selectable between 0 and 2000 liters. 0 volumes are not displayed
Pre-programmed gas types - Air, nitrogen, carbon dioxide, argon
Battery Pack - 20 cells NiMH, 3800mAh, voltage 24 V