Aseptic filling isolators
The main priority is product protection, which is maintained by the double screen principle and Class A cleanroom conditions (ISO 5). A vertical laminar flow of air ensures a constant air speed of 0.45 m/s +/- 20% in the working area. Constant monitoring of viable and non-viable particles as well as of pressure, temperature and humidity takes place. The isolator leak rate is less than 5%. The isolator can be manually wiped down.
Aseptic/toxic filling isolators
The main priority is protecting the product, the operator and the environment. This is achieved by a single screen principle and Class A cleanroom conditions (ISO 5). A vertical laminar flow of air ensures a constant air speed of 0.45 m/s +/- 20% in the working area. Constant monitoring of viable and non-viable particles as well as of pressure, temperature and humidity takes place. The leak rate of the isolator is less than 0.5%. The isolator can be manually wiped down and cleaned without disassembly (CIP).