• User friendly software
• Editing of patient ID, Name (12 characters), Factor
• Print out of laboratory name & Doctor’s name.
• Coagulation test like PT with Magnetic Stirrer.
• Storage of 2000 test results in the memory with patient ID.
• PS2 Keyboard connection optional.
• Lamp Saver mode.
METHODS OF CALCULATIONS
• Absorbance
• End Point with standard / factor
• Kinetics
• Fixed time
• Differential
• Coagulation
• Multipoint upto 10 standards
ASPIRATION SYSTEM
Continuous flow system with bi-directional Pump.
CUVETTES : DUAL MODE
REMOVABLE CUVETTES
Macro, Semi Micro, Micro
Technical Specification
COAGULATION
Mixing by magnetic motor / stirrer INR / ISI Value % / Seconds Auto / Manual start
Parameters - PT / APTT / Fibrinogen
Filter - 405 nm / 630 nm.
ELISA STRIP READER
Wells - 8 position
Filters - 405 nm/450 nm / 630 nm
Methods - Cut-off / Multi standard with EP / KIN / Fix Time
O.D. Range - 0 - 3.5A
OPTICAL SYSTEM
Principle - Absorption spectrometry
Range of measurement - From – 0.1 to 3.0A
Nominal spectral range - Automatic via filter wheel
Filters - 340, 405, 450, 505, 546, 578, 630 and 670nm interference type
Bandwidth - 10 nm (± 2nm)
Reading error - ± 5% at 0.10 A
± 2% at 1.00 A
± 2% at 2.00 A
Light source - Halogen lamp
Zero - Automatic
Settings - Monochromatic and Bichromatic readings.
Detector - Schottkey type Photodiode - Silicon
THERMOST AT SYSTEM
Peltier regulated 25°, 30° & 37°C programmed
Accuracy - ± 0.2°C
CUVETTES - DUAL MODE
Removable cuvettes - Macro, Semi-micro & Micro
PRINTER
Type - Fast Thermal Graphic
Printing speed - 50 mm/sec
Dots per line - 384 dots/line