User friendly software
Editing of patient ID, Name (12 characters), Factor
Print out of laboratory name & Doctor's name.
For Kinetic test : On line Graph & OD. Graph editing facility. Estimation of result.
Universal input voltage 90 to 265 V/50 - 60 Hz.
Storage of 2000 test results in the memory with patient ID.
PS2 Keyboard connection optional. Lamp Saver mode.
METHODS OF ASSAY
ABS
EP with Factor
KIN
Fixed time
Differential
Multipoint upto 10 standards
Technical Specification
OPTICAL SYSTEM
Principle - Absorption spectrometry
Range of measurement - From -0.1 to 3.0A
Normal Spectral range - Automatic via filter wheel
Filters - 340, 405, 505, 546, 578, 630, 670 nm. 1 extra position
Bandwidth - 10nm (±2 nm)
Reading error - 2% at 2.5 A
Light source - Halogen lamp
Zero - Automatic
Settings - Monochromatic & Bichromatic readings.
Detector - Schottkey type photodiode - Silicon
THERMOSTAT SYSTEM
Peltier regulated - 25°, 30° & 37°C
Programmed Accuracy - 0.2°C
CUVETTES - DUAL MODE
Removable cuvettes - Macro, Semi-micro & Micro
PRINTER
Type - Fast Thermal Graphic
Printing speed - 50 mm/sec
Dots per line - 384 dots/line
Paper - Thermo sensitive, 57 mm wide
SCREEN AND KEYBOARD
LCD blue LED back lit - 240 x 64 pixels.
Keyboard - Tactile Soft-touch
ASPIRATION SYSTEM
Continuous flow system with Bi-directional peristaltic pump
INCUBATOR
37°C for 9 Position Test Tubes & 9 Position Cuvettes
PROGRAMMING
Table of units - up to 9 units
Quality control storage of the last 30 results
Memory - 128 KB EEPROM
QUALITY CONTROL
2 Controls per test - Normal & Abnormal