User friendly icon based operation integrated with advanced microprocesor
Calibration cycle: Automatic of programmable: 2hrs / 4 hrs / 6 hrs, standby mode
Liquid stable reagents - RFID reagent management inbuilt
Maintenance free ION SELECTIVE ELECTRODES
Universal input voltage 95 to 265 V / 50-60 Hz
Easy access for electrode assembly
Automatic flagging for low/normal/high results with color code.
Technical Specification
SAMPLE TYPE
Whole Blood / Serum / Plasma / Urine Samples / CSF
SAMPLE VOLUME
120ul Whole Blood, Serum, Plasma
160ul Capillary sample
400ul Diluted (1:4) Urine Sample
SAMPLE ANALYSIS
60 Seconds of Whole Blood, Serum, Plasma / CSF
120 Seconds of Urine Sample
DATA OUTPUT
Ethernet / USB
PRINTER
Thermal Graphic paper 57mm wide
Easy paper loading
SCREEN AND KEYBOARD
Color - Touch Screen 5" : 800 x 478
UNITS
mmol/L or meq/L
DATA STORAGE
1000 test results
Memory: SD card - 8 to 32 GB
QUALITY CONTROL
3 Controls per test - Low, Normal & High
Levy - Jennings charts
Quality control storage of the last 30 results
POWER / DIMENSIONS
Voltage - 110-220 V
Frequency - 50 - 60 Hz
Maximum power - <40W
Temperature - 10° to 35° C
Maximum relative humidity - 80%
Dimensions - 12x12x16 cms
Weight - 8 Kgs